Study on Calibration of Chromlum(VI) Analyzer
Chongqing Academy of Metrology and Quality Inspection, Chongqing 401123, China
a 755245093@qq.com
b 38634546@qq.com
c 718589156@qq.com
d* 755245093@qq.com
e 1020875858@qq.com
This article presents a novel metrological calibration method for trace chromium(VI) analyzer. The study identifies all factors that affect the accuracy of chromium(VI) analyzers, and proposes calibration items and method for the analyzer based on its working principle. The calibration items include pump set flow error and pump flow stability, set temperature error and temperature control stability of column oven, baseline noise and drift, minimum detection concentration, linear range, and qualitative and quantitative repeatability. Experimental results shows that, the presented method can meet the requirements for the metrological performance evaluation of chromium(VI) analyzer. The calibration method is scientific and feasible.
© The Authors, published by EDP Sciences, 2023
This is an Open Access article distributed under the terms of the Creative Commons Attribution License 4.0, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.
Current usage metrics show cumulative count of Article Views (full-text article views including HTML views, PDF and ePub downloads, according to the available data) and Abstracts Views on Vision4Press platform.
Data correspond to usage on the plateform after 2015. The current usage metrics is available 48-96 hours after online publication and is updated daily on week days.
Initial download of the metrics may take a while.