Ionizing radiation metrology at the service of health: quality control of radiopharmaceutical dose calibrators in nuclear medicine unit -accurarcy, reproducibility and linearity tests-
ERSN, Faculty of Sciences, University Abelmalek Essaadi Tetouan, Morocco
National School of Architecture of Marrakech Marrakech, Morocco
National Center for Energy, Nuclear Sciences and Techniques Rabat, Morocco
Bricha National Center for Energy, Nuclear Sciences and Techniques Rabat, Morocco
Laboratory of materials and subatomic physics University ibn tofail Kenitra, Morocco
National Center for Energy, Nuclear Sciences and Techniques Rabat, Morocco
Laboratory of materials subatomic physics University ibn tofail Kenitra, Morocco
Departement of nuclear physics University Abelmalek Essaadi Tetouan, Morocco
Ensuring the effective use of a dose calibrator used for the practice of nuclear medicine, and the injection of the correct dose to the patient is the subject of a long traceable calibration chain, which serves to establish the relationship between the values of the activity indicated by the dose calibrator and the activity corresponding to ISO 2919 certified sealed sources whose activity is known with an uncertainty of less than 1 %.
Our calibration program consists in carrying out various quality control on the CRC 55TR activimeter, namely the daily control which is carried out in the absence of any radioactive source, then the fidelity control carried out by a source of 37Cs and a source of 60Co, then the reproducibility check performed with a 137Cs source, and finally the linearity check performed with a 99mTc source.
In our case the results obtained from the specific tests used to verify the proper functioning of the Capintec CRC 55tR activimeter in a nuclear medicine service in Morocco, shows an agreement with international standards so that the measurement of activities is done with significant results, in addition, to guarantee the metrological quality of the measurements delivered, Our laboratory registers in an accreditation process according to ISO 17025.
Key words: nuclear medicine / quality control / activimeter / ionization / radiation / radiation protection
© The Authors, published by EDP Sciences, 2022
This is an Open Access article distributed under the terms of the Creative Commons Attribution License 4.0, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.
Current usage metrics show cumulative count of Article Views (full-text article views including HTML views, PDF and ePub downloads, according to the available data) and Abstracts Views on Vision4Press platform.
Data correspond to usage on the plateform after 2015. The current usage metrics is available 48-96 hours after online publication and is updated daily on week days.
Initial download of the metrics may take a while.